[pending] CellMek SPS Sample Preparation System, REF: C44603, with software

Pending LLM rewrite. Source: FDA_DEVICE Z-1235-2026.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Pending LLM scoring against the rubric.

Plain-English summary

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

The recalled product

Product
CellMek SPS Sample Preparation System, REF: C44603, with software
Manufacturer
Beckman Coulter, Inc.

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • UDI-DI: 15099590750312. Software Version 2.3.90.0

Distribution

Distributed nationwide across the United States.