The Recall Desk
HighFDA (Devices)·Z-1480-2026·Announced 2026-03-11

[pending] EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Pending LLM rewrite. Source: FDA_DEVICE Z-1480-2026.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: Pending LLM scoring against the rubric.

Plain-English summary

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

The recalled product

Product
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Manufacturer
Staar Surgical AG

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (7)

  • UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964
  • T2286971
  • T2286973
  • T2286979
  • T2286980
  • T2286981
  • T2286988

Distribution

Distribution scope not specified by the agency.