[pending] GEM Premier 5000; Part No. 00055430004.
Pending LLM rewrite. Source: FDA_DEVICE Z-1537-2026.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Pending LLM scoring against the rubric.
Plain-English summary
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
The recalled product
- Product
- GEM Premier 5000; Part No. 00055430004.
- Manufacturer
- Instrumentation Laboratory
- Category
- Medical Device — Devices
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Part number: 00055430004
- UDI: 08426950807711
- Lots No.: 251111O
- 251113U
- 251117AD
- 251118D
- 251119D
- 251120X
- 251124Z
- 251125S
- 251201AI
- 251201AJ
- 251202P
- 251203Q
- 251208AA
- 251212O
- 251215J
- 251216K
- 251217I
- 251217J
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- ModeratePhilips Allura X-ray Systems Deaeration Hose Degradation Recall
FDA (Devices) · 2026-05-27
- HighOncology Kit With Extension Sets May Leak During Infusion Therapy
FDA (Devices) · 2026-05-27
- HighMedtronic GUNDRY Retrograde Cannula vascular catheter sterile barrier breach
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27