[pending] MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electroc
Pending LLM rewrite. Source: FDA_DEVICE Z-0231-2026.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Pending LLM scoring against the rubric.
Plain-English summary
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
The recalled product
- Product
- MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4
- Manufacturer
- GE Medical Systems, LLC
- Category
- Medical Device — Devices
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI: 00840682125499
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- ModerateOneLIF Interbody Cage Inserter Attachment Failure Recall
FDA (Devices) · 2026-05-27
- HighOncology Kit With Extension Sets May Leak During Infusion Therapy
FDA (Devices) · 2026-05-27
- ModeratePhilips Allura X-ray Systems Deaeration Hose Degradation Recall
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Pacing Catheter Model D200F7 Recall Due to Leak Risk
FDA (Devices) · 2026-05-27
- ModerateSilastic Brand Foley Catheters recalled due to surface stain
FDA (Devices) · 2026-05-27