Corin Operating Tables recalled due to electromagnetic interference risk
Corin Operating Tables (6 models) may become unresponsive to commands when used with high-frequency surgical equipment. The issue affects 315 units in the US and worldwide locations.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall involving a risk-of-harm failure mode (surgical equipment unresponsiveness in the operating room). No injuries or adverse events have been reported, so the hazard is documented but not yet manifested in actual harm.
Plain-English summary
MAQUET GMBH is recalling Corin Operating Tables, models 770001A0, 770001A2, 770001B0, 770001B2, 770001F0, and 770001F2. A total of 315 units are affected: 8 distributed in the United States (Kentucky, New York, and Texas) and 307 outside the US in multiple countries.
Under certain conditions, when the operating table is used simultaneously with other medical devices—particularly high-frequency surgical equipment—electromagnetic interference may occur. This interference could prevent the device from responding to commands sent via the universal remote control or override panel, rendering the operating table unresponsive during use.
No injuries or adverse events have been reported as of the recall date. Healthcare facilities using these operating tables should contact the manufacturer for guidance on mitigation measures or device replacement.
The recalled product
- Product
- Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.
- Manufacturer
- MAQUET GMBH
- Category
- Medical Device
- Hazard
- electromagnetic-interference
- device-unresponsiveness
- command-failure
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model Numbers: (1) 770001A0
- (2) 770001A2
- (3) 770001B0
- (4) 770001B2
- (5) 770001F0
- (6) 770001F2
- UDI-DI: (1) 04046768145177
- (2) 04046768145184
- (3) 04046768145191
- (4) 04046768145207
- (5) 04046768145214
- (6) 04046768145221
- Lot/Serial Numbers: (1) 1
- 2
- 3
- 10
- 25
- 26
- 27
- 28
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighGE CARESCAPE Telemetry Server may lose patient ECG and oxygen saturation monitoring
FDA (Devices) · 2026-06-03
- HighPhilips Azurion and Allura Imaging Systems Recalled for Hard Drive Degradation
FDA (Devices) · 2026-06-03
- HighPhilips Allura Imaging Systems Hard Drive Degradation Affects Functionality
FDA (Devices) · 2026-06-03
- ModerateMedtronic Deep Brain Stimulation Pocket Adaptor Kits recalled for incorrect expiration dates
FDA (Devices) · 2026-06-03
- SevereMedline spinal injection kits recalled for quality and efficacy concerns
FDA (Devices) · 2026-06-03