MRI System Helium Venting Malfunction Poses Pressure Leak Risk
Siemens MAGNETOM Skyra MRI systems may have ice blockages in the helium venting system, potentially preventing gas escape during a quench event. This could cause pressure buildup and rupture the helium containment system, releasing helium into the scanning room.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: FDA Class I classification mandates a minimum severity of 4. The hazard involves potential equipment rupture and pressurized gas leak under specific operational conditions (quench event), creating a risk of harm despite no reported injuries to date.
Plain-English summary
Siemens Medical Solutions is recalling 586 MAGNETOM Skyra MRI systems (Model Number 10432915) distributed worldwide. The systems have a potential for ice blockage to form in the magnet venting system.
During a quench event, helium gas is designed to escape safely through designated venting paths. If ice blockage forms or already exists, helium may be unable to escape through these paths. This can lead to pressure buildup within the helium containment system. The pressure buildup could potentially rupture the helium containment system and result in a helium leak into the scanning room.
These devices have been distributed to healthcare facilities worldwide.
The recalled product
- Product
- MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Hazard
- pressure-buildup
- helium-leak
- equipment-rupture
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Model Number: 10432915. UDI Numbers: (01)04056869006703(21)45046
- (01)04056869006703(21)45115
- (01)04056869006703(21)45398
- (01)04056869006703(21)145760
- (01)04056869006703(21)145811
- (01)04056869006703(21)45549
- (01)04056869006703(21)45834
- (01)04056869006703(21)146099
- (01)04056869006703(21)146110
- (01)04056869006703(21)45047
- (01)04056869006703(21)146093
- (01)04056869006703(21)46066
- (01)04056869006703(21)45788
- (01)04056869006703(21)146082
- (01)04056869006703(21)45124
- (01)04056869006703(21)45185
- (01)04056869006703(21)45332
- (01)04056869006703(21)45466
- (01)04056869006703(21)145390
- (01)04056869006703(21)45732
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereMedline spinal anesthesia kits recalled for bupivacaine quality issues
FDA (Devices) · 2026-06-03
- HighGE CARESCAPE Telemetry Server may lose patient ECG and oxygen saturation monitoring
FDA (Devices) · 2026-06-03
- HighGE APEXPRO and CARESCAPE Telemetry Systems Risk Loss of Patient Monitoring
FDA (Devices) · 2026-06-03
- SevereMedline Spinal Trays Recalled Due to Quality and Efficacy Concerns
FDA (Devices) · 2026-06-03
- HighPhilips Azurion and Allura Imaging Systems Recalled for Hard Drive Degradation
FDA (Devices) · 2026-06-03