The Recall Desk

Agency

U.S. Food and Drug Administration

The FDA regulates food, drugs, biologics, medical devices, and tobacco. We aggregate Class I, II, and III enforcement actions across its three recall streams.

Visit the official FDA site →

About FDA recalls

The FDA is the single biggest source of US product recalls by volume. Its enforcement workload splits into three streams — Foods (everything from infant formula to leafy greens to nutritional supplements), Drugs (prescription medicines, over-the-counter products, and active pharmaceutical ingredients), and Medical Devices (everything from implants to home blood-pressure monitors). The Recall Desk ingests all three from the openFDA enforcement endpoint every six hours.

FDA classifies every recall as Class I, II, or III. Class I means there is a reasonable probability that using the product will cause serious health consequences or death — these are the recalls you should treat as urgent. Class II means the product may cause temporary or medically reversible harm. Class III means the product is unlikely to cause adverse health consequences but still violates FDA rules. We map these classifications into our 1–5 severity score using the rubric on the methodology page.

Medical-device recalls are the highest-volume stream by far — software-update letters, calibration corrections, and IFU updates all show up in enforcement reports alongside genuine hazard recalls. We try to surface the severe ones first, but when in doubt, click through to the FDA notice itself for the clinical detail.

16476–16500 of 27263

  • SevereFDA (Devices)·Z-1728-2023·2023-06-28

    Implantable Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic is recalling certain ICD and CRT-D devices that may fail to deliver high-voltage therapy due to a manufacturing defect in the device feedthrough. The recall affects 2,018 units distributed nationwide and worldwide.

    Product
    ICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1746-2023·2023-06-28

    Medtronic Implantable Cardiac Defibrillators Risk Therapy Delivery Failure

    Medtronic is recalling certain CRT-D and ICD models due to a potential feedthrough defect that may prevent energy delivery during high-voltage therapy. The FDA Class I recall affects 435 units distributed in the United States and worldwide.

    Product
    CRT-D DTBB2D4 VIVA S IS1/DF4 INTL, Model Number DTBB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1710-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially compromising treatment effectiveness.

    Product
    ICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1713-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High-Voltage Therapy

    Medtronic implantable cardioverter defibrillators manufactured with a specific feedthrough may experience reduced or no energy output during high-voltage therapy, potentially preventing treatment of life-threatening heart rhythms.

    Product
    ICD-DR DDMB2D1 EVERA MRI XT OUS DF1, Model Number DDMB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1720-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver High Voltage Therapy

    Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during high voltage therapy due to a specific feedthrough manufacturing defect. Approximately 8,219 units were affected worldwide.

    Product
    ICD-DR DDME3D4 MIRRO MRI, Model Number DDME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1734-2023·2023-06-28

    Medtronic Implantable Cardioverter Defibrillators Recalled for Reduced Therapy Output

    Medtronic is recalling 14 units of CRTD VIVA QUAD XT implantable cardioverter defibrillators due to rare potential for reduced or no energy output during high voltage therapy. Affected devices were manufactured with a specific feedthrough defect.

    Product
    CRTD VIVA QUAD XT CUST PARY GOLD, Model Number DTBA1QQPG; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1806-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail to Deliver Therapy

    Medtronic is recalling 15,381 implantable cardioverter defibrillators due to rare potential for reduced or absent energy output during high-voltage therapy caused by a specific feedthrough design defect.

    Product
    ICD-VR VISIA MRI AF XT OUS DF4, Model Number DVFB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1753-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High-Voltage Therapy Energy

    Certain Medtronic implantable cardioverter defibrillators may produce reduced or no energy output during therapy due to a feedthrough defect. Approximately 11,794 units worldwide are affected.

    Product
    CRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1813-2023·2023-06-28

    Implantable Defibrillators Risk Therapy Failure Due to Feedthrough Defect

    Medtronic implantable defibrillators may fail to deliver therapy during critical cardiac events due to a feedthrough defect. Approximately 25,058 units are affected nationwide and worldwide.

    Product
    ICD-VR DVMC3D4 EVERA MRI S DF4 GLOB, Model Number DVMC3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1799-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Energy Output Failure

    Medtronic's implantable cardioverter defibrillators may fail to deliver high-voltage therapy due to a feedthrough defect. Approximately 6,821 devices are affected worldwide.

    Product
    ICD-VR DVBC3D1 EVERA S IS1/DF1 GLOB, Model Number DVBC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1727-2023·2023-06-28

    Implantable cardioverter defibrillators may output reduced or no energy during therapy

    Medtronic ICDs and CRT-Ds with specific feedthroughs may fail to deliver adequate high-voltage therapy. This FDA Class I recall affects 487 units distributed nationwide and worldwide.

    Product
    ICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1758-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver Therapeutic Shock

    Medtronic implantable cardioverter defibrillators with a specific feedthrough may produce reduced or no energy output during high-voltage therapy, potentially failing to deliver life-saving shocks.

    Product
    CRTD DTMA2D4 CLARIA MRI OUS DF4, Model Number DTMA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1783-2023·2023-06-28

    Implantable Defibrillator Models May Fail to Deliver Shock Therapy

    Medtronic implantable cardiac defibrillators may fail to deliver electrical therapy during treatment due to a defective feedthrough component. Approximately 1,878 devices are affected.

    Product
    CRTD COBALT HF QUAD MRI IS4 DF1, Model Number DTPB2Q1 ; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1706-2023·2023-06-28

    Implantable Defibrillators Recalled for Potential Therapy Failure During High-Voltage Treatment

    Medtronic is recalling specific Implantable Cardioverter Defibrillator (ICD) models due to a rare potential for reduced or no energy output during high-voltage therapy. The issue is related to a specific manufacturing defect in the device feedthrough.

    Product
    ICD EVERA XT DR GOLD CTD, Model Number DDBB1D4G; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1714-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential Reduced Energy Output

    Medtronic implantable cardioverter defibrillators (ICDs) and CRT-Ds with specific feedthrough designs may produce reduced or insufficient energy during high voltage therapy, affecting the device's ability to treat heart arrhythmias.

    Product
    ICD-DR DDMB2D4 EVERA MRI XT IS-1/DF4 INT, Model Number DDMB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1747-2023·2023-06-28

    Implantable Cardioverter Defibrillators Recalled for Potential High-Voltage Therapy Failure

    Medtronic is recalling 7 implantable cardioverter-defibrillators that may fail to deliver electrical therapy during critical heart rhythms. The defect affects devices with a specific feedthrough component.

    Product
    CRT-D DTBB2Q1 VIVA QUAD S IS4/DF1 INTL, Model Number DTBB2Q1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1798-2023·2023-06-28

    Implantable Cardioverter Defibrillators May Fail To Deliver Therapy; Class I Recall

    Medtronic implantable cardioverter defibrillators (ICDs) may fail to deliver high-voltage therapy due to a manufacturing defect in a specific feedthrough component. 447 units are affected worldwide.

    Product
    ICD-VR DVBB2D4 EVERA XT DF4 INTL, Model Number DVBB2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1776-2023·2023-06-28

    Medtronic Implantable Defibrillators May Fail Therapy Energy Delivery

    Medtronic recalls 3 COBALT XT implantable defibrillators that may deliver reduced or no energy during therapy. A defect in the device feedthrough could prevent effective treatment of abnormal heart rhythms.

    Product
    CRTD DTPA2D1PX COBALT XT HF QUAD OUS, Model Number DTPA2D1PX; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1704-2023·2023-06-28

    Implantable defibrillators may fail to deliver high-voltage therapy due to defect

    Medtronic's EVERA XT implantable defibrillators may fail to deliver high-voltage therapy in certain models manufactured with a defective feedthrough. Patients should consult their cardiologist.

    Product
    ICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1721-2023·2023-06-28

    Implantable Defibrillators May Fail to Deliver High Voltage Therapy

    Certain Medtronic implantable cardioverter defibrillators may deliver reduced or no energy during high voltage therapy due to a defective feedthrough. This defect could prevent the device from properly treating life-threatening heart rhythms.

    Product
    ICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1737-2023·2023-06-28

    Medtronic Implantable Heart Defibrillators Risk Therapy Delivery Failure

    Medtronic CRT-D and ICD implantable defibrillators may fail to deliver therapy due to a rare defect in a specific feedthrough component. The recalled devices carry a potential for reduced or no-energy output during high-voltage heart rhythm treatment.

    Product
    CRT-D DTBA2D4 VIVA XT IS1/DF4 INTL, Model Number DTBA2D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1711-2023·2023-06-28

    Medtronic Cardiac Defibrillators May Fail to Deliver Therapy

    Certain Medtronic implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) may produce reduced or no energy output during high voltage therapy due to a defective feedthrough component.

    Product
    ICD-DR DDMB1D1 EVERA MRI XT US DF1, Model Number DDMB1D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1817-2023·2023-06-28

    Implantable Cardiac Defibrillators Recalled for Potential Therapy Delivery Failure

    Medtronic's ICD-VR DVME3D4 defibrillators are recalled due to a rare risk of reduced or no energy output during high-voltage therapy. The defect involves a specific feedthrough component and affects approximately 14,989 units distributed nationwide and worldwide.

    Product
    ICD-VR DVME3D4 MIRRO MRI, Model Number DVME3D4; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1793-2023·2023-06-28

    Implantable Defibrillators May Produce Reduced or No Energy Output

    Medtronic implantable cardioverter defibrillators may produce reduced or no-energy output during high voltage therapy due to a defective feedthrough component. The defect affects 176 units distributed worldwide.

    Product
    ICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • SevereFDA (Devices)·Z-1805-2023·2023-06-28

    Implantable Cardioverter Defibrillators at Risk of Reduced Energy Output

    Medtronic ICDs and CRT-Ds with a specific feedthrough defect may produce reduced or no energy output during high-voltage therapy. The recall involves approximately 4,546 units distributed nationwide and worldwide.

    Product
    ICD-VR DVFB2D1 VISIA AF MRI XT OUS DF1, Model Number DVFB2D1; Implantable Cardioverter Defibrillators
    Category
    Medical Device
    Distribution
    Distributed nationwide