The Recall Desk

Agency

U.S. Food and Drug Administration

The FDA regulates food, drugs, biologics, medical devices, and tobacco. We aggregate Class I, II, and III enforcement actions across its three recall streams.

Visit the official FDA site →

About FDA recalls

The FDA is the single biggest source of US product recalls by volume. Its enforcement workload splits into three streams — Foods (everything from infant formula to leafy greens to nutritional supplements), Drugs (prescription medicines, over-the-counter products, and active pharmaceutical ingredients), and Medical Devices (everything from implants to home blood-pressure monitors). The Recall Desk ingests all three from the openFDA enforcement endpoint every six hours.

FDA classifies every recall as Class I, II, or III. Class I means there is a reasonable probability that using the product will cause serious health consequences or death — these are the recalls you should treat as urgent. Class II means the product may cause temporary or medically reversible harm. Class III means the product is unlikely to cause adverse health consequences but still violates FDA rules. We map these classifications into our 1–5 severity score using the rubric on the methodology page.

Medical-device recalls are the highest-volume stream by far — software-update letters, calibration corrections, and IFU updates all show up in enforcement reports alongside genuine hazard recalls. We try to surface the severe ones first, but when in doubt, click through to the FDA notice itself for the clinical detail.

16676–16700 of 27263

  • HighFDA (Devices)·Z-1687-2023·2023-06-14

    Avanos Medical Recalls ambIT Reusable Cassettes Missing Air-Inline Filter

    Avanos Medical is recalling SMMT ambIT Reusable cassettes (Product Code 220568, Lot 30191805) that were potentially distributed without a required air-in-line filter. The missing component affects units in Texas, Oregon, and Australia.

    Product
    SMMT ambIT Reusable Program - Luer Cass, Product Code: 220568
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1676-2023·2023-06-14

    Diagnostic Test Recall: 4Kscore Test Software Anomalies May Produce Erroneous Results

    BioReference Health is recalling the 4Kscore Test system due to software anomalies in its processing module that may generate erroneous diagnostic results. The recall affects 662 patient test results distributed in New Jersey.

    Product
    The 4Kscore Test is an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1686-2023·2023-06-14

    SMMT ambIT Kit Basic Filter Luer Cass Recalled for Missing In-Line Filter

    Avanos Medical recalls SMMT ambIT Kit Basic Filter Luer Cass units that were distributed without the required air in-line filter component.

    Product
    SMMT ambIT Kit Basic Filter Luer Cass, Product Code: 220537
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1678-2023·2023-06-14

    DePuy Mitek BIOKNOTLESS Plus Anchor Inserter Tip May Fracture During Surgery

    DePuy Mitek BIOKNOTLESS Plus anchor inserter tips may break and be retained in patients during surgery. Removal of fragments could cause bone damage and prolong surgery time.

    Product
    DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1680-2023·2023-06-14

    DePuy Mitek surgical anchor inserter may fracture during use

    DePuy Mitek BIOKNOTLESS Rapid surgical anchor inserters may fracture during use, potentially retaining fragments in patients. Removal during surgery may cause bone damage and extended surgical time.

    Product
    DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0997-2023·2023-06-14

    Chili Cacahuate Japones Snacks Recalled for Undeclared Peanut

    Werner Gourmet Meat Snacks recalls Chili Cacahuate Japones snacks due to undeclared peanut allergen. The product poses a serious risk to consumers with peanut allergies.

    Product
    Chili Cacahuate Japones, net weight 5 oz (127g), packaged in a soft teal colored plastic pouch with the UPC code 7 09972 80082 2. There are 6 pouches per case. Werner Gourmet Meat Snacks Inc. Tillamook, OR 97141. The label declares: INGREDIENTS: CHILI, RICE FLOUR, WHEAT FLOUR
    Category
    Food
    Distribution
    13 states
  • HighFDA (Food)·F-0994-2023·2023-06-14

    Children's colostrum supplement recalled due to undeclared milk allergen

    Nature's Energy Children's Chewable Colostrum Berry supplement is recalled because the label does not declare milk allergen, despite containing bovine colostrum. The recall affects consumers with milk allergies.

    Product
    Nature's Energy Children's Chewable Colostrum Berry, 180 count, UPC 804879183945 GTIN 860008676979
    Category
    Food
    Distribution
    0 states
  • HighFDA (Devices)·Z-1685-2023·2023-06-14

    AVNS ambIT System infusion kits recalled for missing air filter

    Avanos Medical is recalling certain AVNS ambIT System infusion kits because some units may lack the required air in-line filter. Affected kits were distributed worldwide including the US (Texas and Oregon) and Australia.

    Product
    AVNS ambIT System - PIB/PCA W 5in Cath, Product Code: 220507
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1677-2023·2023-06-14

    Surgical Anchor Inserter Tip Fragmentation During Patient Procedures

    DePuy Mitek's BIOKNOTLESS Plus anchor inserter may fragment during surgical use if bending force is applied. Fragments retained in the patient may require surgical removal, potentially causing bone damage.

    Product
    DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1675-2023·2023-06-14

    Elekta Unity Image-Guided Radiation Therapy System Electrical Connector Overheating Risk

    Elekta, Inc. is recalling 18 Elekta Unity Image-Guided Radiation Therapy Systems due to a potential electrical connector overheat hazard in the MR gradient coil. No injuries have been reported.

    Product
    Elekta Unity Image-Guided Radiation Therapy System containing an MR gradient coil.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1683-2023·2023-06-14

    QIAstat-Dx Respiratory SARS-CoV-2 Test Kit Recalled for False Result Risk

    Qiagen Sciences LLC recalls 347 QIAstat-Dx Respiratory SARS-CoV-2 test kits due to faulty cartridges that could produce false results. Affected kits were distributed nationwide in nine states.

    Product
    QIAstat-Dx Respiratory SARS-CoV-2 Panel multiplexed nucleic acid real-time PCR test, Reference Number 691223
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1679-2023·2023-06-14

    DePuy Mitek surgical anchor inserter tip may fragment and be retained

    DePuy Mitek BIOKNOTLESS Plus surgical anchor inserters may have tips that break off during use and remain in the patient. This could require surgical removal that may cause bone damage.

    Product
    DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1682-2023·2023-06-14

    BD Veritor System Instructions for Use Not FDA Approved

    The Instructions for Use for BD Veritor System rapid diagnostic tests for COVID-19 and flu were not FDA-approved, lacking proper specimen type specifications. Using the test with invalid specimen types could produce false positive or negative results, potentially delaying diagnosis or treatment.

    Product
    BD Veritor System for Rapid Detection of SARS-CoV-2 & Flu A+B, 30 Single Use Test Devices- Intended for the in vitro, simultaneous qualitative detection and differentiation of SARS-CoV-2 nucleocapsid antigen and influenza A and/or B Catalog Number: 256088
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1691-2023·2023-06-14

    Sterile Surgical Procedure Packs Recalled for Defective Light Handle Covers

    DeRoyal Industries is recalling 2,972 sterile surgical procedure packs containing light handle covers that have been recalled by another manufacturer. The affected packs were distributed to hospitals in Virginia.

    Product
    Sterile custom surgical procedure packs: (1) 89-10507.02, TOTAL KNEE PACK, Stonewall Jackson Hospital; (2) 89-10508.02, LOCAL PACK, Stonewall Jackson Hospital; (3) 89-10509.02, LAPAROTOMY PACK, Stonewall Jackson Hospital; (4) 89-10517.04, CNRV BASIC SETUP PACK, New River Va
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1681-2023·2023-06-14

    Surgical anchor inserter recalled for potential tip fracture and bone damage

    DePuy Mitek BIOKNOTLESS Rapid anchor inserters may fracture during use, potentially leaving fragments in the patient. If detected during surgery, removal could cause bone damage.

    Product
    DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Food)·F-0972-2023·2023-06-14

    Blueberry Yogurt Parfaits Recalled for Undeclared Soy and Brown Sugar

    Sodexo Operations recalls simply to go Blueberry Yogurt Parfaits due to undeclared soy lecithin and brown sugar syrup. Consumers allergic to soy should not consume the product.

    Product
    simply to go Blueberry Yogurt Parfaits, 8.53oz, UPC 477088156552 and 477088150550; packaged in plastic cups
    Category
    Food
    Distribution
    3 states
  • HighFDA (Devices)·Z-1698-2023·2023-06-14

    Magic3 Go Female Intermittent Urinary Catheters Recalled for Packaging Defect

    C.R. Bard Inc is recalling Magic3 Go female intermittent urinary catheters worldwide due to a manufacturing defect where the resealing label can separate from the product pouch, potentially compromising product sterility.

    Product
    Magic3 Go Intermittent Urinary Catheters, Female Catalog Numbers: a) REF 51808, b) REF 51810, c) REF 51812, d) REF 51814, e) REF 51816, f) REF 51818
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Food)·F-0996-2023·2023-06-14

    Nature's Energy Children's Chewable Colostrum Recalled for Undeclared Milk Allergen

    Nature's Energy is recalling Children's Chewable Colostrum Orange Creamsicle (180 count) due to undeclared milk allergen on the label. Consumers with milk allergies could experience serious allergic reactions.

    Product
    Nature's Energy Children's Chewable Colostrum Orange Creamsicle, 180 count, UPC 804879183969 GTIN 860008676993
    Category
    Food
    Distribution
    0 states
  • HighFDA (Devices)·Z-1699-2023·2023-06-14

    Magic3 Go Intermittent Urinary Catheters Recalled for Packaging Defect

    C.R. Bard is recalling Magic3 Go Intermittent Urinary Catheters because the clear resealing label can separate from the pouch, potentially compromising sterility. The recall affects devices distributed worldwide.

    Product
    Magic3 Go Intermittent Urinary Catheters, male Catalog Numbers: a) REF 50810G, b) REF 50812G, c) REF 50814G, d) REF 50816G, e) REF 50818G, f) REF 53810G, g) REF 53812G, h) REF 53814G, i) REF 53816G, j) REF 53818G, k) REF 53820G
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Food)·F-0981-2023·2023-06-14

    Chocolate spread recalled for undeclared soy allergen

    DIBE LLC is recalling TETAS Mireya, SABOR Nutella chocolate spread due to undeclared soy allergen. Consumers with soy allergies should not consume this product.

    Product
    TETAS Mireya, SABOR Nutella
    Category
    Food
    Distribution
    0 states
  • HighFDA (Drugs)·D-0865-2023·2023-06-14

    Empower Pharmacy Ascorbic Acid Injection Mislabeled as Preserved

    Empower Pharmacy is recalling Ascorbic Acid 500 mg/mL injection solution because vials labeled as preserved are actually preservative-free. The 504 affected vials were distributed nationwide.

    Product
    ASCORBIC ACID PRESERVED INJECTION SOLUTION, 500 mg/mL, 30 mL Sterile Multiple-Dose Vial, RX ONLY, Compounded by: Empower Pharmacy, 5980 W Sam Houston Pkwy N Ste 300 Houston, TX 77041, NDC 72627-2405-1
    Category
    Drug
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-1690-2023·2023-06-14

    Presource Laparoscopy Kit Recalled for Lack of Sterility Assurance

    Cardinal Health is recalling Presource Kit laparoscopy surgical packs due to being released without sterility assurance. Affected units may harbor contamination and pose a risk during surgical procedures.

    Product
    Presource Kit, SMA21LPUHJ GYN Laparoscopy Pack. Convenience kit used for Laparoscopic surgery
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1689-2023·2023-06-14

    TREO Abdominal Stent-Graft Label Does Not Match Actual Device Size

    Bolton Medical is recalling the TREO Abdominal Stent-Graft System because the carton label may not match the actual stent-graft size inside. Implanting the wrong size stent-graft could result in surgical complications.

    Product
    The TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graft and a Straight Iliac Extension Stent-Graft. Each stent-graft is preloaded into its own delivery s
    Category
    Medical Device
    Distribution
    Distributed nationwide