Small intestinal videoscope recalled due to inadequate channel drying process
Aizu Olympus is recalling small intestinal videoscopes (Model SIF-Q180) because the channel air drying process was not properly validated. Water remaining in the channels could cause contamination and infection risk.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a risk-of-harm medical device recall where the potential for infection exists due to contamination from inadequate channel drying validation, but no illnesses or injuries have been reported. The theoretical hazard falls under the rubric criterion for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Aizu Olympus Co., Ltd. is recalling the small intestinal videoscope Model SIF-Q180, a medical instrument used for endoscopy and endoscopic surgery of the digestive tract. The recall affects 32 units that have been distributed worldwide, including throughout the United States.
The company determined that the channel air drying process was not properly validated during the repair procedure. A small percentage of scopes returned to customers after repair were found to have water remaining in the channels. Water in the channels creates a risk of contamination that could lead to infection.
The recalled product
- Product
- Small intestinal videoscope-For endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach, duodenum, and small intestine) by oral insertion or the lower digestive tract (including the anus, rectum, sigmoid colon, colon, ileocecal valve, a
- Manufacturer
- Aizu Olympus Co., Ltd.
- Hazard
- contamination
- infection-risk
- inadequate-drying
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Model Number (UDI-DI): SIF-Q180 (04953170339936)
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27