Radiation therapy planning software displays wrong images during animation playback
RayStation/RayPlan versions 6.0.0.24 through 9.2.0.483 incorrectly display selected images as both primary and secondary when animation playback is running, affecting treatment planning visualization.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device with a display defect affecting radiation therapy treatment planning visualization. No illnesses or injuries reported; this is a preventable risk-of-harm scenario addressed through software updates.
Plain-English summary
RayStation/RayPlan is FDA-cleared medical device software designed for radiation therapy treatment planning and analysis. Multiple software versions (6.0.0.24 through 9.2.0.483) contain a defect in the image display system.
When a user selects a new primary image set while the cine loop (animation playback) is running, the software incorrectly displays the selected image set as both the primary and secondary image in all side-by-side viewing modes. The same display error occurs when a new patient or case is opened while the cine loop is running.
The defect affects radiation therapy centers across the United States where 75 licenses of the software are deployed. Users should contact RaySearch Laboratories AB for available software updates and guidance to prevent treatment planning errors.
The recalled product
- Product
- RayStation/RayPlan- designed for treatment planning and analysis of radiation therapy. Software Version: 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 . Includi
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Hazard
- display-error
- treatment-planning-defect
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (13)
- Software: 6.0.0.24
- 6.1.0.26
- 6.1.1.2
- 6.2.0.7
- 6.3.0.6
- 7.0.0.19
- 8.0.0.61
- 8.0.1.10
- 8.1.0.47
- 8.1.1.8
- 8.1.2.5
- 9.0.0.113
- 9.1.0.933
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- ModerateZeiss Intrabeam Spherical Applicator failed manufacturing specifications
FDA (Devices) · 2026-07-08
- SevereOmnipod 5 Insulin Infusion Pods Recalled for External Cannula Damage
FDA (Devices) · 2026-07-08
- HighBoston Scientific CRE Pro Wireguided 12-15mm Esophageal Stent Recall
FDA (Devices) · 2026-07-08
- HighMedline Convenience Kits with Recalled Swan-Ganz Catheters Nationwide
FDA (Devices) · 2026-07-08
- HighMedline Convenience Kits with Recalled Swan-Ganz Catheters
FDA (Devices) · 2026-07-08