HeartWare HVAD Pump Kit Driveline Cover Hardening Issue
HeartWare HVAD Pump Kit driveline covers may harden over time, making it difficult or impossible to access the driveline connector for urgent servicing. Delayed access could result in patient harm from prolonged pump stoppage.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a Class II device recall with no reported illnesses or injuries. The hazard is a risk-of-harm product (potential delayed access to driveline connector in a life-sustaining cardiac device) where injury has not yet been reported, meeting the High-level rubric criterion.
Plain-English summary
The HeartWare HVAD Pump Kit, REF 1104 (UDI/DI 00888707006323), is subject to this recall. Approximately 10,320 units have been distributed worldwide.
Over time, the driveline cover on the pump kit may become hardened. This hardening creates difficulty or an inability to slide the cover back to access the driveline-to-controller connector.
If urgent access to the driveline connector is needed to address a driveline connection or controller issue, a hardened cover could delay necessary servicing. This delay may result in patient harm from prolonged pump stoppage.
Patients and healthcare providers who observe a hardened driveline cover should contact their Medtronic field representative for further assessment and servicing.
The recalled product
- Product
- HeartWare HVAD Pump Kit, REF 1104
- Manufacturer
- Heartware, Inc.
- Hazard
- material-hardening
- connector-access-obstruction
- delayed-service-risk
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- UDI/DI 00888707006323
- Serial Numbers: HW25347
- HW26099
- HW26126
- HW26274
- HW26281
- HW26330
- HW26332
- HW26660
- HW26762
- HW26939
- HW26940
- HW26969
- HW27069
- HW27099
- HW27154
- HW27184
- HW27596
- HW27674
- HW27678
Distribution
Distribution scope not specified by the agency.
Related recalls
Same category
- HighMedtronic DLP One-Piece Pediatric Arterial Cannula Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- SevereT-Lok Bone Marrow Tray Kits Recalled for Sterility Testing Failure
FDA (Devices) · 2026-07-22
- SevereMedtronic Bio-Medicus Adult Cannula Kit Recalled for Foreign Material
FDA (Devices) · 2026-07-22
- SevereAbiomed Impella CP Smart Assist Set Heart Pump Introducer Sheath Leakage
FDA (Devices) · 2026-07-22