Intuitive Surgical Clip Applier Recalled for Frayed or Broken Grip Cables
Intuitive Surgical is recalling 5,576 units of the 8MM Horizon Small Clip Applier due to complaints of frayed or broken grip cables on the reusable instruments.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Class II recall with no reported injuries or hospitalizations. Grip cable failures on a surgical instrument represent a risk-of-harm defect without reported injury, meeting the High (3) severity criterion.
Plain-English summary
Intuitive Surgical, Inc. is recalling 5,576 units of the 8MM Horizon Small Clip Applier (Model 470401) distributed nationwide in the United States and internationally. The reusable surgical instrument comes in two part numbers: 470401-07 and 470401-08.
The manufacturer received increased complaints regarding frayed or broken grip cables on the instrument jaws. Multiple batch numbers manufactured between 2023 and 2024 are affected, including batch numbers K10230129 through K11240112.
Healthcare facilities and providers using affected devices should identify units by batch number and contact Intuitive Surgical, Inc. for instructions regarding handling of recalled devices.
The recalled product
- Product
- 8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
- Manufacturer
- Intuitive Surgical, Inc.
- Hazard
- cable-fraying
- cable-breakage
Distribution
Distribution scope not specified by the agency.
Related recalls
Same category
- HighMedtronic DLP Retrograde Cannula cardiopulmonary bypass catheter
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27
- HighPhilips Azurion X-Ray Systems Table Movement Control Defect Recall
FDA (Devices) · 2026-05-27