VOCSN Pediatric breathing package spiral wrap detachment risk during ventilation
Ventec Life Systems recalls VOCSN Patient Breathing Packages because the spiral wrap may detach during ventilation, compromising the breathing circuit and creating a risk of serious injury or death, especially for ventilator-dependent patients.
What this means for you
Highest-severity recall. Risk of serious injury or death is documented or strongly suspected. Stop using the product now and follow the agency's instructions.
Our severity reasoning: FDA Class I medical device classification combined with a confirmed manufacturing defect affecting a life-critical respiratory device. The hazard is not theoretical—spiral wrap detachment is occurring—and poses explicit risk of serious injury or death, particularly in ventilator-dependent patients.
Plain-English summary
Ventec Life Systems, Inc. is recalling 250 units of the VOCSN Patient Breathing Package (Pediatric, Active, Oxygen) with reference numbers PRT-01013-000 (package) and PRT-00802-001 (pouch). The product is used with the VOCSN Unified Respiratory System to provide continuous or intermittent ventilatory support for patients requiring mechanical ventilation.
The bonded spiral wrap in the breathing circuit may detach before or during ventilation due to a manufacturing issue. Detachment can compromise the structural integrity, functionality, and performance of the device, resulting in occlusions (blockages) or leaks in the breathing circuit.
This defect creates a serious risk of injury or death, particularly for ventilator-dependent patients who rely on continuous respiratory support. The affected products were distributed in the United States to facilities in California, Colorado, Florida, Hawaii, Indiana, Missouri, North Carolina, and Washington, as well as to Japan. Affected units have lot number 220922 or pack slip lot number 257337.
Patients and healthcare providers should immediately cease use of affected units and contact Ventec Life Systems for instructions on replacement or repair.
The recalled product
- Product
- VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ven
- Manufacturer
- Ventec Life Systems, Inc.
- Hazard
- detachment
- obstruction
- leak
- malfunction
Distribution
Distributed in 8 states:
- CA
- CO
- FL
- HI
- IN
- MO
- NC
- WA
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27