Surgical Convenience Kits Recalled Due to Compromised Sterility of External Wrapping
Medline Industries is recalling 630 surgical convenience kits because the perforated bag design fails to maintain sterility of the outer wrapping, though internal contents remain sterile.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is a theoretical risk-of-harm situation involving medical device sterility without reported illnesses or injuries. The Class II FDA classification indicates a potentially serious condition, but the absence of any reported harm limits the severity to the High category.
Plain-English summary
Medline Industries, LP is recalling 630 units of surgical convenience kits distributed nationwide. The recalled products include the Medline General Utility kit (Reorder Number DYNDA2524), Centurion Medical Products Dialysis On/Off Bundle (Reorder Number DT22705A), Centurion Medical Products Ultrasound PIV Bundle (Reorder Number IV8635D), and Centurion Medical Products Cutdown Tray (Reorder Number MNS12595).
The recall is due to a design flaw in the perforated bag that prevents the outside of the kit wrapping from remaining sterile. Only the contents inside the kit wrapping maintain sterile integrity. This compromised outer sterility could potentially allow contamination during storage or transport before the kit is used.
Affected lot numbers are: DYNDA2524 (Lot 22HME893), DT22705A (Lot 22HMC897), IV8635D (Lot 22HMH318), and MNS12595 (Lot 22HMD598). Healthcare facilities and providers who have received these kits should follow instructions provided by Medline Industries regarding the recall.
The recalled product
- Product
- Surgical convenience kits labeled as: a) MEDLINE GENERAL UTILITY, Reorder Number DYNDA2524 b) CENTURION MEDICAL PRODUCTS DIALYSIS ON/OFF BUNDLE, Reorder Number DT22705A c) CENTURION MEDICAL PRODUCTS ULTRASOUND PIV BUNDLE, Reorder Number IV8635D d) CENTURION MEDICAL PROD
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Category
- Medical Device — Surgical Kits
- Hazard
- sterility-compromise
- contamination-risk
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (9)
- a) Reorder Number DYNDA2524
- GTIN 40193489361415
- Lot # 22HME893 b) Reorder Number DT22705A
- GTIN 40653160993224
- Lot # 22HMC897 c) Reorder Number IV8635D
- GTIN 40653160990261
- Lot# 22HMH318 d) Reorder Number MNS12595
- GTIN 40653160991015
- Lot # 22HMD598
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighPhilips Avalon Fetal Monitor FM 30 Speaker Connector Assembly
FDA (Devices) · 2026-07-01
- SevereDa Vinci surgical system operative components with E-brake retainer pin defect
FDA (Devices) · 2026-07-01
- ModerateOctopus Evolution AS Tissue Stabilizer Model TS2500 Tubing Assembly Issue Recall
FDA (Devices) · 2026-07-01
- ModerateOctopus 4 Tissue Stabilizer incorrectly assembled tubing positions
FDA (Devices) · 2026-07-01
- ModerateMedline Polycarbonate Syringes in Convenience Kits Recalled for Unapproved Design Changes
FDA (Devices) · 2026-07-01