HeartWare HVAD Battery Model 1650DE May Malfunction Due to Weld Defect
Medtronic is recalling HeartWare HVAD batteries (Model 1650DE) due to an internal weld defect that can cause battery malfunction and prevent proper charging. Approximately 429 affected batteries were distributed worldwide.
What this means for you
Illness, injury, or structural failure has been reported. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: This recall is classified as FDA Class I, which requires a minimum severity score of 4. A weld defect affecting a battery component in a ventricular assist device creates risk of malfunction that could disrupt continuous power delivery to a life-critical cardiac device.
Plain-English summary
Medtronic Inc. is recalling HeartWare HVAD batteries, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) system. A total of 429 batteries with specific serial numbers are affected, carrying GTIN 00888707008778 with an expiration date of April 30, 2022.
The recalled batteries contain a weld defect affecting internal components. This defect can cause the battery to malfunction, resulting in failure to provide power or inability to hold a complete charge or recharge properly. The HVAD is a ventricular assist device used to support heart function in patients with advanced heart disease.
The affected batteries were distributed worldwide, including throughout the United States and in Austria, Belgium, Finland, France, Germany, Hungary, Italy, Netherlands, New Zealand, North Macedonia, Norway, Poland, Spain, Switzerland, and the United Kingdom.
Patients or healthcare providers with affected HVAD systems should verify their battery serial number against the recall list and contact Medtronic or their healthcare provider to determine appropriate next steps. Do not discontinue device use without consulting a physician.
The recalled product
- Product
- HeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.
- Manufacturer
- Medtronic Inc
- Hazard
- weld-defect
- battery-malfunction
- power-failure
- charging-failure
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- GTIN 00888707008778. Serial numbers BAT935012
- BAT935059
- BAT935008
- BAT935009
- BAT935018
- BAT935020
- BAT935022
- BAT935023
- BAT935047
- BAT934999
- BAT935003
- BAT935010
- BAT935329
- BAT934928
- BAT934934
- BAT935160
- BAT935078
- BAT934997
- BAT935027
- BAT935048
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighGE Healthcare AW Server Radiological Image Processing System Software Defect
FDA (Devices) · 2026-07-01
- HighSurgify Halo Surgical Burr Recalled for Breakage Risk
FDA (Devices) · 2026-07-01
- ModerateThuvera Laser Console Software Version 2.2 Recall Notice
FDA (Devices) · 2026-07-01
- ModerateMedline Polycarbonate Syringes Recalled for Unapproved Design Changes
FDA (Devices) · 2026-07-01
- SevereMedline Polycarbonate Syringes in Convenience Kits Recalled for Unapproved Design Changes
FDA (Devices) · 2026-07-01