Roadrunner PC Wire Guide Recalled for Sterility Testing Failure
Cook Incorporated is recalling 1,997 units of the Roadrunner PC Hydrophilic Wire Guide due to packaging failure in sterility testing. The device's sterility may be compromised.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall for potential sterility compromise due to packaging failure. With no reported illnesses or injuries in the source text and a potential rather than confirmed hazard, this meets the threshold for High (score 3) rather than Severe (score 4) per the rubric.
Plain-English summary
Cook Incorporated is recalling 1,997 units of the Roadrunner PC Hydrophilic Wire Guide (Reference Part Number RPC-35-180), a device used for vascular catheter positioning and exchange in diagnostic and interventional procedures. The recall was initiated because the product did not meet acceptance criteria for packaging integrity testing, which included 3-year age acceleration testing. As a result, the sterile packaging may have deteriorated, potentially compromising the device's sterility.
The affected devices were distributed nationwide and internationally, including multiple countries across Europe, Asia, Middle East, Africa, and the Americas. The recall affects devices with lot numbers 14063361, 14087089, 14087092, 14117033, 14120926, 14120928, 14151216, 14151218, 14153159, 14158614, 14159695, and 14159696.
The recalled product
- Product
- Roadrunner PC Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): RPC-35-180 ORDER NUMBER (GPN):G07518. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
- Manufacturer
- Cook Incorporated
- Hazard
- sterility-compromise
- infection-risk
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (12)
- UDI-DI: (01)00827002075184 Lot Numbers: 14063361
- 14087089
- 14087092
- 14117033
- 14120926
- 14120928
- 14151216
- 14151218
- 14153159
- 14158614
- 14159695
- 14159696
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27