Zimmer Distal Lateral Fibular Plate may fail to lock properly
Zimmer orthopedic plates may have thread defects preventing locking screws from properly mating, potentially resulting in loss of fixation and need for additional surgery.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device recall involving a manufacturing defect in locking mechanisms that could result in loss of surgical fixation requiring re-operation. No injuries or hospitalizations have been reported; the hazard is potential rather than actualized, meeting the rubric criterion for High severity: 'risk-of-harm products where injury has not yet been reported.'
Plain-English summary
Zimmer, Inc. has recalled ZPLP Distal Lateral Fibular Plates (Item Number 7-2357-018-04) used in orthopedic surgery. The affected lots contain a thread form defect in the locking holes that may prevent locking screws from properly mating with the plate.
An improperly mated screw may not lock correctly, and this defect may not be immediately recognizable to the surgeon. This could result in loss of fixation during surgery, potentially requiring additional surgical intervention.
Affected units have been distributed worldwide, including to the United States, Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, South Korea, Singapore, Switzerland, and Taiwan. The 14 affected lot numbers are detailed in the recall notice.
Healthcare providers should verify whether devices in their inventory match the affected lot numbers and discontinue use of affected units. Patients who have received this implant should contact their surgeon if they experience complications.
The recalled product
- Product
- ZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 7-2357-018-04
- Manufacturer
- Zimmer, Inc.
- Hazard
- fixation-loss
- screw-locking-defect
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27