X-ray table recalled for fire hazard from power supply unit short circuit
Siemens MULTIX TOP I PRO X-ray tables are recalled due to a power supply short circuit that may cause overheating and fire. Three units have been identified.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall with a potential fire hazard, but no reported injuries, illnesses, property damage, or actual incidents. The hazard is theoretical ('may lead to') rather than confirmed, fitting the criterion for score 3: risk-of-harm products where injury has not yet been reported.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling the MULTIX TOP I PRO Radiographic X-ray Table (Material Number 475525). The device is a patient table used in radiographic imaging systems in conjunction with Siemens Multix PRO and TOP Systems devices. Three units with serial numbers 1523, 1585, and 1897 are affected.
A short circuit has been identified in the power supply unit (PSU) of the Touch Display component. This short circuit may lead to overheating and potential fire.
The affected units have been distributed nationwide in the United States. Facilities that have received these devices should contact Siemens Medical Solutions USA, Inc. to determine appropriate corrective action for units with the identified serial numbers.
The recalled product
- Product
- MULTIX TOP - Intended Use: The Multix TOP I PRO Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 475525
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Hazard
- fire
- electrical-short-circuit
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI-DI: N/A Serial Numbers: 1523 1585 1897
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27