Landauer nanoDot radiation dosimeters recalled for accuracy drift
Landauer is recalling nanoDot D2DNS radiation dosimeters worldwide because some units may read outside the specified ±5.5% accuracy range due to a potential non-conformance in the Optical Stimulated Luminescence material.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall affecting a risk-of-harm product (radiation dosimeter) where accuracy is critical to function. No illnesses, injuries, or hospitalizations have been reported. Per the rubric, risk-of-harm products without reported injury are High (3), not Severe.
Plain-English summary
Landauer is recalling nanoDot D2DNS radiation monitoring dosimeters (Model 03066-OTO), which are used with microSTAR readers. All batch numbers distributed worldwide are affected.
Investigation found that some dosimeters may operate outside the specified accuracy range of ±5.5%. The issue is a potential non-conformance in the Optical Stimulated Luminescence (OSL) material used in the dosimeter. The OSL material is stimulated by beaming after radiation exposure and emits fluorescence proportional to the radiation dose received.
The recalled product
- Product
- QC (80 kVp) screened nanoDot D2DNS, Model Number 03066-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Manufacturer
- Landauer
- Hazard
- accuracy-drift
- osl-defect
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- UDI/DI 0860003399903
- all batch numbers
Distribution
Distribution scope not specified by the agency.
Related recalls
Same category
- SevereMedtronic DLP Retrograde Cannula recalled for sterile barrier breach risk
FDA (Devices) · 2026-05-27
- HighAzurion X-ray System Deaeration Hose May Degrade and Cause Oil Leakage
FDA (Devices) · 2026-05-27
- HighSwan-Ganz Jr Catheter models recalled due to lumen hub leakage
FDA (Devices) · 2026-05-27
- SevereMedtronic DLP Retrograde Cannula Recalls for Sterile Barrier Breach Risk
FDA (Devices) · 2026-05-27
- HighMedtronic DLP Retrograde Cannula Cardiopulmonary Bypass Catheter Recall
FDA (Devices) · 2026-05-27