Permobil M3 Corpus Power Wheelchair Recalled Over Firmware Stability Issue
Permobil is recalling M3 Corpus Power Wheelchairs with firmware version 1.26.2 because the firmware lacks a speed restriction when the seat lift exceeds 50 mm, which may cause falls or tip-overs.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The FDA classified this as Class II, and the source does not report any injuries or hospitalizations. The hazard is a potential fall or tip-over without reported harm, matching the High criterion for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Permobil Inc. is retroactively recalling the Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock, REF 120031, Model version 3. The recall covers all units operating with firmware version 1.26.2 (UDI (01)17330818345674). A total of 140 units were distributed worldwide, including nationwide distribution in the United States and in Canada.
The firmware in these wheelchairs lacks a speed restriction needed to maintain stability when the seat lift exceeds 50 mm. This condition may cause the user to fall from the wheelchair or the wheelchair to tip over.
The recall applies to users of the affected wheelchairs with firmware version 1.26.2. The U.S. Food and Drug Administration has classified this as a Class II recall. The recall notice does not specify next steps for consumers.
The recalled product
- Product
- Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
- Manufacturer
- Permobil Inc
- Hazard
- fall
- tip-over
- firmware-defect
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All Units operating with firmware version 1.26.2/UDI:(01)17330818345674
Distribution
Distributed nationwide across the United States.
Related recalls
Same category
- HighMedical Convenience Kits Recalled Over Endotoxin Variability in Neurological Sponges
FDA (Devices) · 2026-08-05
- HighStryker Hoffmann II Carbon Connecting Rods Recalled Over Potential Cracking After Sterilization
FDA (Devices) · 2026-08-05
- HighPeripherally Inserted Central Catheters Recalled Due to Hub Leaking and Cracking Risk
FDA (Devices) · 2026-08-05
- HighBlood Analyzer Bar Code Reader May Assign Patient Results to Incorrect Patient ID
FDA (Devices) · 2026-08-05
- ModerateMedical Convenience Kits Recalled Over Lidocaine Supplier Quality Issue
FDA (Devices) · 2026-08-05