The Recall Desk

Recalls

All US Product Recalls

Searchable archive of product recalls from the FDA, USDA FSIS, CPSC, NHTSA, and EPA — translated into plain English, severity-scored from objective criteria, and always linked back to the original government notice.

How to read this listing

Every recall on this list has been scored 1–5 against an objective rubric from the agency’s own classification, presence of reported injuries or deaths, distribution scope, and hazard type. The colored severity badge on each card is the score — 5 (Critical) means deaths or serious injuries are reported, 4 (Severe) means significant injury or hospitalization risk, and 3 (High) means a known hazard with no confirmed harm yet. Filter to 3+ (High and above) if you want to skip routine Class-III notices.

The agency chips above control which streams are visible. FDA spans food, drugs, and medical devices. USDA FSIS is meat, poultry, and processed egg products. CPSC is household consumer products. NHTSA is vehicles, tires, and child seats. Nationwide on a card means the recall affects every US state — most do — while a state list means distribution was limited.

Click any recall to see the full summary, affected lot codes and model numbers, the agency’s own classification, and a direct link back to the original government notice for verification.

Filters

Agency

Category

Minimum severity

Clear filters

4976–5000 of 13449

  • HighFDA (Devices)·Z-3141-2024·2024-09-25

    Alaris Syringe Module Model 8110 Infusion Pumps Recalled for Bent Connectors

    CareFusion's Alaris Syringe Module Model 8110 infusion pumps are being recalled because affected units may have bent female connectors that can prevent device operation and delay infusions. Approximately 20,145 units have been distributed worldwide.

    Product
    Alaris Syringe Module Model 8110, a modular infusion pump and monitoring system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3171-2024·2024-09-25

    K-Wire Packaging Error Recall by Smith & Nephew

    Smith & Nephew issued a voluntary recall of K-Wire bone fixation fasteners due to a packaging error where packages contained the wrong wire diameter than labeled.

    Product
    DIAMOND POINT, BOTH ENDS K-WIRE, REF 128042, metallic bone fixation fastener
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-3150-2024·2024-09-25

    Sump Catheter System Recalled for Potential Sterile Barrier Defect

    Boston Scientific is recalling the vanSonnenberg Sump Catheter System Kit because the sterile barrier pouch may have holes that could compromise device sterility. The recall affects 4 units distributed in the US, Colombia, Philippines, and Russia.

    Product
    vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System Kit, Material Number REF M001273090; for use as a percutaneously placed drain for intra-abdominal fluid collections.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-3165-2024·2024-09-25

    Philips Azurion X-ray Systems: Cable Hose Carrier Detachment Risk

    Philips recalls Azurion X-ray imaging systems due to potential cable hose carrier detachment from broken plastic or bolts, which may cause parts to fall and create injury risk.

    Product
    System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azu
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3152-2024·2024-09-25

    Medtronic CareLink 2090 Programmer Recall: Autonomous Cursor Motion

    Medtronic is recalling CareLink 2090 Programmers (serial number prefixes PKK0 and PKK1) worldwide due to potential autonomous cursor motion when Finger Touchscreen capability is enabled with software version 3.2 or higher. Approximately 18,788 units are affected.

    Product
    2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090 PROGRAMMER-PORTABLE CRM; b) 209031, Product Description: PROGRAMMER 209031 PORTABLE CRM; c) 2090AA, Product Description: PROGRAMMER 2090AA PORTABLE CRM; d) 2090AA4, Pr
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-3157-2024·2024-09-25

    Medical Thrombectomy Device Recalled for Incorrect Expiration Dates

    MicroVention is recalling 1,370 ERIC Retrieval Devices due to labeling errors showing incorrect expiration dates (5 years instead of actual 3 years). Devices with mislabeled shelf life may be used beyond their true expiration.

    Product
    ERIC Retrieval Device REF: ER173020 ER174030 ER176044 The ERIC Retrieval Device is a mechanical thrombectomy device designed to restore blood flow by effectively removing clots from occlusive vasculature in patients suffering from acute ischemic stroke.
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3143-2024·2024-09-25

    Seer Home EEG System Recalled for EMC Non-Compliance

    Seer Home System ambulatory EEG devices were recalled due to electromagnetic compatibility non-compliance and manufacturing control deficiencies. The issue may affect signal recording reliability in home EEG monitoring.

    Product
    Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory comp
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3172-2024·2024-09-25

    Smith & Nephew K-Wire Surgical Device Recall for Packaging Error

    Smith & Nephew is recalling K-Wire surgical devices due to a packaging error that resulted in wrong-sized wires being distributed. Healthcare providers may have received incorrectly sized bone fixation fasteners affecting approximately 73 units.

    Product
    DIAMOND POINT, BOTH ENDS K-WIRE, REF 128062, metallic bone fixation fastener
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-1157-2024·2024-09-25

    Medline Contro-Bulb Syringe Recalled for Potential Sterile Barrier Breach

    Medline is recalling 4,831 Contro-Bulb Syringes due to potential sterile barrier breach during transportation. The recall affects two lot numbers and applies nationwide.

    Product
    Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3149-2024·2024-09-25

    Nephrostomy Catheter System Kit Recalled Due to Sterile Barrier Defect

    Boston Scientific is recalling Flexima Regular Kit Nephrostomy Catheter System kits due to potential holes in the sterile barrier pouch that may compromise device sterility.

    Product
    Flexima Regular Kit Nephrostomy Catheter System Kit, Material Numbers REF M001271860; To provide external drainage of the urinary tract.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-3151-2024·2024-09-25

    Boston Scientific VTC Nephrostomy Catheter Kit Recalled for Sterile Barrier Defect

    Boston Scientific is recalling VTC Regular Kit Nephrostomy Catheter System kits due to potential holes in the sterile barrier of the device pouch, which may compromise device sterility.

    Product
    VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the urinary tract.
    Category
    Medical Device
    Distribution
    0 states
  • HighFDA (Devices)·Z-3173-2024·2024-09-25

    Hemodialysis Bloodline Tubing Recall Over Incomplete Connector Insertion Risk

    B Braun is recalling SL BLOODLINE FOR FMC 2008 SERIES hemodialysis tubing nationwide due to incomplete connector insertion that may cause blood flow restrictions during treatment.

    Product
    SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3175-2024·2024-09-25

    Hemodialysis bloodline tubing recalled for improper connector insertion risk

    B. Braun is recalling approximately 4.98 million STREAMLINE FRESENIUS FOR DAVITA hemodialysis bloodline tubes due to risk of improper connector insertion, which could cause blood flow occlusion during dialysis treatment.

    Product
    STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095DA
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3180-2024·2024-09-25

    Beckman Coulter Analyzer Recalled for Pneumatic Tubing Defects

    Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized pneumatic tubing in the reagent storage module, which can cause air leaks and compressor failures, potentially leading to test errors and delayed results.

    Product
    Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137; in vitro diagnostic instrument
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3140-2024·2024-09-25

    Alaris EtCO2 Module Model 8300 Recalled Due to Bent Connector

    CareFusion 303, Inc. recalls 180 Alaris EtCO2 Module Model 8300 units with potentially bent female connectors that may prevent device operation, risking delays in infusion and monitoring.

    Product
    Alaris EtCO2 Module Model 8300, a modular infusion pump and monitoring system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3164-2024·2024-09-25

    Endo Retrieval Pouches Recalled; Tube May Detach During Endoscopic Use

    Endo retrieval pouches are recalled because the tube may detach and fall into the surgical site during use. Approximately 81,241 units distributed in Georgia, Missouri, and Pennsylvania are affected.

    Product
    1. Tech Medical Services Inc. Detachable Endo Retrieval Pouch, REF TM2001, 3" x 6", sterile; 2. Tech Medical Services Inc. Detachable Endo Retrieval Pouch, REF TM2000, 5" x 7", sterile; 3. Tech Medical Services, Inc. Detachable Endo Retrieval Pouch, REF TM2002, 8" x 10", s
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3181-2024·2024-09-25

    HAWK Warming Grid recalled for incorrect expiration date

    Safeguard US Operating LLC is recalling the HAWK Warming Grid (Item 59-300) due to an incorrect expiration date. The affected units may not function safely if used beyond the actual expiration date.

    Product
    HAWK Warming Grid, Item Number: 59-300
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3155-2024·2024-09-25

    Shoulder implant recalled due to excessive bacterial endotoxin

    Tornier, Inc. is recalling one lot of the Stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS shoulder implant due to bacterial endotoxin exceeding safety specifications. The 5 affected units were distributed in Illinois, Wisconsin, Arizona, Nevada, and South Dakota.

    Product
    stryker TORNIER PERFORM HUMERAL SYSTEM - STEMLESS US Anatomic Nucleus, REF DWM003A; anatomic total arthroplasty of the shoulder
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3161-2024·2024-09-25

    BiomarC EX Fiducial Marker System recalled for missing MRI safety component

    Carbon Medical Technologies is recalling the BiomarC EX Fiducial Marker System because patients did not receive the required MRI Insert component needed for safe use during MRI procedures.

    Product
    BiomarC EX Fiducial Marker System, Catalog Number 040165
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3142-2024·2024-09-25

    Alaris Pump Module 8100 Recalled for Bent Connector That May Disable Device

    CareFusion is recalling approximately 192,964 Alaris Pump Module Model 8100 infusion pumps that may have bent female connectors preventing device operation and delaying patient treatment.

    Product
    Alaris Pump Module Model 8100, a modular infusion pump and monitoring system
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3166-2024·2024-09-25

    Philips Allura Xper Fluoroscopy Systems Cable Hose Carrier Detachment

    Philips Allura Xper FD series fluoroscopy systems may have detached cable hose carriers due to broken plastic or bolts. A detachment could cause parts to fall and potentially result in injury.

    Product
    System Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR Table 722038 Allura Xper FD20/20 722035 Allura Xper FD20 OR Table 722029 Allura Xper FD20/10 722028 Allura Xper FD20 722027 Allura Xper FD10/10 722026 Allura Xper FD10 722025
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3144-2024·2024-09-25

    Anesthesia Extension Sets recalled due to potential Spin Lock connector leakage

    B. Braun Medical is recalling 19,050 AET36 Anesthesia Extension Sets due to potential product leakage from the Spin Lock Connector caused by excess solvent migration during manufacturing.

    Product
    AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3159-2024·2024-09-25

    Convenience kits recalled for containing defective medical syringes

    ROi CPS LLC is recalling medical convenience kits distributed with Merit syringes that were subsequently recalled by their manufacturer. Approximately 3,387 kits with multiple lot numbers were affected.

    Product
    regard Clinical Packaging Solutions, HT00376I - SPECIAL PROCEDURE, Item Number 800223009
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3168-2024·2024-09-25

    Philips MultiDiagnost Eleva X-ray system cable hose carrier detachment

    Philips is recalling 109 MultiDiagnost Eleva X-ray systems worldwide due to potential cable hose carrier detachment from broken plastic or bolts, which could cause parts to fall and result in injury.

    Product
    System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 708036 MultiDiagnost Eleva 708034 MultiDiagnost Eleva with Flat Detector 708032 MultiDiagnost Eleva
    Category
    Medical Device
    Distribution
    Distributed nationwide
  • HighFDA (Devices)·Z-3176-2024·2024-09-25

    Hemodialysis bloodline connector insertion defect may cause flow restriction

    B Braun Medical is recalling hemodialysis bloodline tubing because the patient connector may not fully insert into the device, potentially blocking blood flow and delaying treatment. Affected units were distributed nationwide.

    Product
    STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095L
    Category
    Medical Device
    Distribution
    Distributed nationwide